Medical Equipment Regulations (Registration of Medical Equipment in the Register and Renewal thereof), 5773-2013

תקנות ציוד רפואי (רישום ציוד רפואי בפנקס וחידושו), תשע"ג-2013

Published: 2013-06-12Consolidated Hebrew text as of 2023-09-18 · Last amended 2022-05-10
Premium
Unofficial English translation — for reference only. It may contain errors or omissions and cannot be relied on as a legal text. Only the Hebrew text published in Reshumot is legally binding.More

This English text was translated from the official Hebrew using a range of translation tools, and it undergoes ongoing checks and updates. It is not a certified translation.

Despite these checks, it may contain errors, omissions, or imprecise renderings of legal terminology and cross-references, and it may not yet reflect the latest amendments. It cannot be relied upon as a legal text.

The Hebrew text as published in Reshumot (ספר החוקים) and on the Knesset website is the sole authoritative and legally binding version. In any discrepancy, the Hebrew text prevails.

This translation is provided for informational purposes only and does not constitute legal advice. For use in legal proceedings, request a certified Expert Legal Opinion.

By virtue of my authority under Sections 3(c) and (d), 6(c), 8(e) and 22 of the Medical Equipment Law, 5772-2012 (hereinafter – the Law), I hereby make these Regulations:

Definitions§

1.

In these Regulations –

"qualified professional" means a resident of Israel whom the applicant has appointed in writing as a contact person pursuant to regulation 21, and who meets the training requirements set out in regulation 22;

"applicant" means a person submitting an application for the registration of medical equipment in the Register, or a registration holder submitting an application for the amendment or renewal of the registration of the medical equipment;

"marketer" means a person engaged in marketing as defined in the Law;

"low-risk non-sterile medical equipment" means medical equipment appearing on the list of non-sterile medical equipment of the Class 1 Non Sterile type, appearing in the publication of the British health authority (MHRA) as updated from time to time;

"preparation" has the meaning assigned to it in the Pharmacists Ordinance [New Version], 5741-1981;

"ISO standard" means a standard of the International Standard Organization;

"ISO 9001 standard" means an ISO standard for external quality assurance in design, development, manufacture, installation, provision of service and maintenance;

"ISO 13485 standard" means an ISO standard for quality assurance in the manufacture of medical equipment.

Submission of Application by a Contact Person or Qualified Professional§

2.

An application for the registration of medical equipment in the Register, for the amendment of a registration or for the renewal of a registration shall be submitted to the Director by a resident of Israel whom the applicant has appointed for that purpose as referred to in regulation 21 (hereinafter – contact person) or by a qualified professional, as the case may be.

Application for Registration in the Register§

3.
(a)An application for registration in the Register shall be submitted in an electronic format as the Director shall publish and shall include the following particulars and documents:
(1)particulars of the business in Israel, its address and the particulars of the contact person or qualified professional, as well as the business licence number and the date of expiry of the licence;
(2)a certificate as detailed in regulation 20, attesting that the storage and transportation of the medical equipment comply with the appropriate storage and transportation conditions;
(3)particulars of the manufacturer of the medical equipment and the manufacturing site, and a certificate attesting that the medical equipment is manufactured under appropriate manufacturing conditions as required by regulation 19;
(4)the full name of the medical equipment that is the subject of the application for registration, in Hebrew and in English, the full particulars of the models for which approval is sought, a physical description of the equipment, as well as a description of its technical, medical and functional characteristics;
(5)a description of the indication and the intended use of the equipment and the purpose of use for which registration in the Register is sought;
(6)the names of the distributors and sub-suppliers of the medical equipment through whom the medical equipment is imported;
(7)the presence of a preparation, chemical substance, poison or biological substance in the medical equipment, as well as an indication of whether the equipment emits ionising or non-ionising radiation;
(8)certificates attesting that the equipment is registered in a foreign state, as well as an official certificate of that state confirming the actual marketing of the equipment within its territory, if any;
(9)particulars regarding the manner in which support and maintenance services are provided for the medical equipment and the identity of the service providers;
(10)particulars regarding the labelling and marking of the equipment and the packaging, as well as a proposed package insert, to the extent required by the Director;
(11)proof of payment of a fee pursuant to Section 22 of the Law for the submission of an application for registration in the Register;
(12)declarations by the applicant and the manufacturer of the medical equipment regarding the provision of complete and reliable information on the medical equipment as required under the Law and these Regulations;
(13)an application submitted by a qualified professional shall be accompanied by the documents confirming the training required under regulation 22.
(b)In an application for the registration of medical equipment that is registered in a recognised state or is permitted for marketing therein, this shall be stated explicitly, and the applicant shall also state the registration periods approved in the recognised state, the conditions of registration or restrictions imposed therein, if any, as well as instructions issued regarding the labelling and packaging of the equipment.
(c)An application for the registration of medical equipment manufactured in Israel that is not registered or permitted for marketing in a recognised state shall also be accompanied by the results of a risk management survey (Risk Analysis) conducted by the applicant in respect of the equipment, a clinical evaluation (Clinical Evaluation), a file containing a summary of the clinical trials conducted on the equipment, as well as two professional opinions from medical experts in the relevant field regarding the safety and efficacy aspects of the medical equipment.
(d)An application for the registration of medical equipment that is substantially equivalent to medical equipment registered in the Register in Israel or in a recognised state shall be submitted as referred to in sub-regulations (a) and (b); for this purpose, "equivalent medical equipment" means medical equipment intended for the same use as the registered medical equipment, having the same technological characteristics, and whose safety and efficacy profile is identical to the profile of the registered equipment.
(e)Where the application for registration concerns medical equipment intended for use in a health institution only, the applicant shall state this explicitly.

Request for Additional Information and Sample§

4.

The Director may require an applicant to provide documents or additional information not included in the application, as well as samples of the medical equipment, and may put questions to the applicant on any matter relating to the registration of the medical equipment.

Making a Decision on an Application§

5.
(a)The Director shall communicate his decision on an application for registration of the equipment within 120 days of the date of submission of the application, and if the application is refused, shall communicate his decision in a reasoned written notice.
(b)Where the Director has required an applicant to complete data or to provide additional data, the period until the receipt of such data shall be added to the time period prescribed in sub-regulation (a).
(c)Where the Director has approved an application for the registration of medical equipment in the Register, he shall issue a certificate of registration of the equipment in the Register within three working days of the date on which he communicated his decision.

Submission of an Application for Amendment of an Existing Registration§

6.
(a)An application for the amendment of a registration in the Register shall be submitted in an electronic format as the Director shall publish and shall include the particulars and documents detailed in regulation 3(a), with the necessary modifications.
(b)The provisions of regulation 5(a) to (c) shall apply to the Director's decision on an application for amendment of a registration in the Register.

Submission of an Application for Renewal§

7.
(a)An application for the renewal of a registration in the Register shall be submitted in an electronic format as the Director shall publish and shall include the particulars detailed in regulation 3(a)(1) and (3); the application shall include the remaining particulars in regulation 3, to the extent that any change has occurred therein since the original application for registration was submitted.
(b)An application for the renewal of a registration in the Register shall be accompanied by a summary of the follow-up and audit activities carried out by the registration holder pursuant to Section 11 of the Law, in a format as directed by the Director; a health institution submitting an application for the renewal of a registration as aforesaid shall attach the summary of the follow-up and audit activities carried out at the health institution.
(c)The Director may require that an applicant for the renewal of a registration attach to the application a professional opinion from a professional who uses the medical equipment in Israel, regarding the use of the registered medical equipment.

Giving Notice of Continued Marketing of Low-Risk Non-Sterile Equipment§

8.
(a)A registration holder of low-risk non-sterile medical equipment registered in the Register shall send to the Director, 30 days before the expiry of the validity of the registration of the medical equipment, a notice of continued marketing of medical equipment in the format that the Director shall publish on the Ministry's website, if no change whatsoever has occurred in the registration particulars and documents submitted in the original application; the notice shall include the particulars detailed in regulation 3(a)(1) and (3) as well as the summary of the follow-up and audit activities carried out by the registration holder pursuant to Section 11 of the Law, in the format as directed by the Director on the Ministry's website, and shall also be accompanied by a receipt for payment of the fee pursuant to Section 22 of the Law for the submission of a notice of renewal of registration in the Register.
(b)Where a notice of continued marketing as referred to in sub-regulation (a) has been submitted in time, it shall be accompanied by a declaration by the registration holder that no special event connected with the use of the medical equipment has occurred, as detailed in Section 11(a)(3) of the Law, and that no market withdrawal activities have been carried out; where such a notice has been submitted and the Director has not notified his objection to the continued marketing within fourteen working days of the date of its submission, the registration shall be deemed to have been extended for a period of five years, and the registration holder may request from the Director a certificate attesting that the validity of the registration of the medical equipment has been extended pursuant to the notice of the registration holder; such a certificate shall be issued by the Director to the registration holder within three working days of the date of receipt of the request.

Special Provisions regarding a Combination Product§

9.
(a)Where an application for the registration of medical equipment incorporating a preparation has been submitted, the Director shall decide whether it is to be registered in the medical equipment Register as a combination product or whether the preparation is to be registered in the register of medicinal preparations maintained pursuant to Section 47a of the Pharmacists Ordinance [New Version], 5741-1981.
(b)Before the Director decides on the classification of the product as a preparation or as medical equipment, he shall examine its primary mode of action and primary use (Primary Mode of Action), and shall also examine what expertise is required of the person applying for the registration of the product.
(c)For the purpose of the classification of the product, the Director may require the applicant to provide additional documents and data necessary for making the decision.
(d)A preliminary application for the classification of a product as medical equipment or as a preparation shall be submitted to the Medical Equipment Department of the Ministry of Health; a decision on the application shall be given within 30 working days of the date of its submission.

Priority of Equipment Classification by the Health Authority of the United States§

10.

Where medical equipment is registered in more than one recognised state, including with the health authority of the United States of America (FDA), the Director shall adopt the classification as well as the conditions of registration and registration periods as determined by the health authority of the United States of America.

Rejection of Applications Submitted Incompletely§

11.
(a)The Director shall not accept an application for the registration of medical equipment in the Register or an application for the renewal or amendment of a registration as aforesaid, and shall not process such an application, unless it has been submitted in full together with all required documents and forms; where an application has been submitted incompletely, the Director shall reject it without further reasons.
(b)Where an application has been rejected as referred to in sub-regulation (a), the fee paid in respect thereof shall not be refunded and re-submission of the same application shall not be permitted until one month has elapsed from the date on which notice of its rejection was given; where such an application has been submitted two additional times and rejected due to incomplete submission as referred to in sub-regulation (a), the applicant shall not be able to submit it again unless the Director has approved this in a reasoned written decision.
(c)Notwithstanding the provisions of sub-regulation (b), where a fee was paid in respect of an application that was rejected due to incomplete submission as referred to in sub-regulation (a), no additional fee shall be charged for the re-submission of the same application within the six months following the giving of notice of its rejection.

Period of Registration of Medical Equipment in the Register§

12.

Where the Director has approved the registration of medical equipment in the Register, he shall register it for a period not exceeding five years; however, medical equipment of the type detailed below shall be registered for the periods indicated alongside it:

(1)implants – three years;
(2)medical equipment registered or permitted for marketing in a recognised state – until the expiry of the validity of the registration or permit in that state, provided that the registration period shall not exceed 5 years from the date of registration in Israel.

Market Withdrawal§

13.

Where the Director or the registration holder considers that medical equipment does not meet the registration requirements, or that it is defective in a manner liable to cause harm to a patient, its marketing shall be immediately ceased, and the registration holder shall take the measures necessary for the withdrawal of the medical equipment from the market in accordance with the Director's instructions.

Non-Renewal of Registration of Medical Equipment§

14.

Where a registration holder does not apply for the renewal of the registration of medical equipment, he shall notify the Director thereof in writing before the expiry of the validity of the registration; in his notice, the registration holder shall state the reason for the non-renewal and whether sufficient substitutes for the registered equipment exist on the market; the registration holder shall publish a notice to health institutions and to the public regarding the non-renewal of the registration of the medical equipment and the cessation of its marketing, and a copy of the notice shall be sent to the Director.

Notice of Changes§

15.
(a)Where a registration holder becomes aware at any time, including after the registration of medical equipment in the Register, of a change in any matter from among those listed in regulation 3(a), the registration holder shall notify the Director and the health institution that purchased the equipment thereof as soon as possible, through the contact person or the qualified professional, as the case may be.
(b)Regarding an intention to make a change in a particular from among the particulars listed in regulation 3, a registration holder shall submit an application for approval of the change pursuant to regulation 6.

Issuance of a Copy of a Registration Certificate§

16.

At the request of a registration holder, the Director shall issue to him an official copy of a certificate of registration of medical equipment in the Register, within seven working days of the date on which he submitted his request.

Import Approval§

17.

Where the Director has registered medical equipment in the Register, he shall give the registration holder or his authorised agent an import approval for the medical equipment into Israel.

Certificate of Free Sale in Israel§

18.

The Director may, at the request of a registration holder, issue to him a certificate attesting that specific medical equipment is registered in the Register and is permitted for marketing in Israel.

Appropriate Manufacturing Conditions for Medical Equipment§

19.
(a)Medical equipment marketed in Israel shall be manufactured at a plant holding a certificate from a health authority in the state of manufacture or any other competent body in the state of manufacture, according to which the appropriate manufacturing conditions at the plant comply with the requirements of ISO standard 13485, including amendments made thereto from time to time.
(b)Medical equipment manufactured in Israel shall not be marketed unless the manufacturing plant holds an approval from a body recognised by the Director, confirming that the appropriate manufacturing conditions therein comply with the requirements of ISO standard 13485, as well as a business licence for the manufacture of medical equipment pursuant to the Business Licensing Law, 5728-1968 (hereinafter – the Licensing Law).

Appropriate Storage and Transportation Conditions§

20.

Medical equipment marketed in Israel shall be transported and stored by a person holding an approval from a body recognised by the Director, confirming that his transportation conditions comply with the requirements of ISO standard 9001, as well as a business licence for the storage and transportation of medical equipment pursuant to the Licensing Law.

Contact Person§

21.

An applicant shall appoint from among its employees a contact person who shall be responsible for submitting applications under these Regulations and responding to enquiries from the Director concerning medical equipment registered in the Register; the appointment of the contact person shall be made in writing and a copy thereof shall be sent to the Director and also to any medical institution purchasing the medical equipment from the applicant.

Training Requirements for a Qualified Professional§

22.

In respect of applications for the registration of medical equipment of the types specified in paragraphs (1) to (3), the contact person shall be a qualified professional holding the training set out alongside the types of applications below:

(1)in respect of an application for the registration of life-supporting and life-sustaining medical equipment and implants – a physician, or a holder of a recognised first degree in life sciences, or a holder of a recognised first degree in medical sciences, or a holder of a biomedical engineer's certificate recognised by the person authorised to manage the Register of Engineers and Architects under the Engineers and Architects Law, 5718-1958 (hereinafter – the Registrar of Engineers);
(2)in respect of an application for the registration of medical equipment that incorporates a preparation, equipment intended for the treatment of blood products, and medical equipment containing a substance of animal origin – a qualified professional as referred to in paragraph (1), or a licensed pharmacist as defined in the Pharmacists Ordinance [New Version], 5741-1981;
(3)in respect of an application for the registration of imaging devices, equipment that emits ionising or non-ionising radiation, and equipment using laser technology – a qualified professional as referred to in paragraph (1), or a holder of an electronics engineer's certificate recognised by the Registrar of Engineers, or a holder of a first degree in physics.

Labelling and Marking of Medical Equipment§

23.
(a)No person shall market medical equipment in Israel unless the following particulars are indicated on the packaging of the equipment or in a leaflet contained in the packaging, as the Director shall decide, in the Hebrew, Arabic and English languages, in a legible, clear and indelible manner:
(1)the trade name of the medical equipment;
(2)the intended use of the medical equipment and its indications;
(3)the name of the manufacturer and the country of manufacture of the medical equipment;
(4)the name and address of the registration holder of the equipment and a telephone number for responding to public enquiries;
(5)warnings and restrictions on the use of the medical equipment as the Director shall direct, and additional information relating to the safe use of the equipment; where a leaflet setting out the warnings and information as aforesaid is enclosed with the packaging, there shall be indicated on the packaging, in a conspicuous manner, an instruction in the following wording: "The enclosed leaflet should be read before use";
(6)the batch number or serial number of the equipment and the date of manufacture in the format directed by the Director;
(7)the expiry date, if any, in the format directed by the Director;
(8)the quantity of items in the packaging;
(9)where the equipment also contains a preparation, the active substances of the preparation and their quantities in a dose of the preparation shall be set out, by their generic names in Latin characters;
(10)any other marking directed by the Director in the registration certificate of the medical equipment.
(b)No medical equipment shall be marketed in Israel unless the batch numbers of the equipment and the serial number, if any, the date of manufacture and any other marking indicated by the Director in the registration certificate appear on the equipment itself, in a legible, clear and indelible manner.
(c)The Director may, at the request of a distributor, exempt the distributor from the requirement of markings under sub-regulation (a), whether all or some of them, in the three languages as aforesaid, on medical equipment not intended for use by a domestic user, if the Director is of the opinion that this would not impair the safety of use thereof.

Notice of Identity of Distributors§

24.

A registration holder shall notify the Director, at the Director's request, of the identity of the distributors to whom the registration holder has transferred the medical equipment, and shall also retain the documents relating to the identity of the distributors as aforesaid for a period of at least three years from the date on which a batch was transferred to them.

Commencement§

25.

These Regulations shall come into force 30 days from the date of their publication.

Read the entire law on one page — continuous text, no page breaks, plus formal citations.

RegulationsHealthCorporate & BusinessCivil Law

תקנות ציוד רפואי (רישום ציוד רפואי בפנקס וחידושו), תשע"ג-2013

Medical Equipment Regulations 2013